TB 9-6625-133-40
TI
Range (Ω)
Mode
Cal Point
Nominal
Test Report
20 M
Normal
Zero
Short
------
10 M Ω
20 M
Normal
Range Gain
20 M
LoI
Zero
Short
------
10 M Ω
20 M
LoI
Range Gain
HiVΩ
20 M
Zero
Short
------
HiVΩ
10 M Ω
20 M
Range Gain
1
200 M
Normal
Zero
Short
------
100 M Ω
1
200 M
Normal
Range Gain
1
200 M
LoI
Zero
Short
------
100 M Ω
200 M1
LoI
Range Gain
HiVΩ
1
200 M
Zero
Short
------
HiVΩ
100 M Ω
1
200 M
Range Gain
1
2G
Normal
Zero
Short
------
1GΩ
2 G1
Normal
Range Gain
2 G1
LoI
Zero
Short
------
1GΩ
1
2G
LoI
Range Gain
HiVΩ
1,2
2G
Zero
Short
------
HiVΩ
1GΩ
2 G1,2
Range Gain
Disable 4-wire Ω mode for all measurements > 20 MΩ.
1
200 M Ω and 2 G Ω Normal Ω must be adjusted before 2 G HiVΩ for the result to be valid.
2
(13) Repeat technique of (9) through (14) above for remaining TI ranges and modes
that require adjustment.
(14) Exit CAL menu by pressing [QUIT] 2 times.
(15) Disconnect equipment set-up.
e. DC Current Adjustments
NOTE
Just one range may be adjusted if required, but for a full
adjustment, start with the 200 μA range and work up to the
20 A range. Zero cal point on all ranges requiring adjustment
may be completed before any range gain adjustments to avoid
multiple connect/disconnects. Any range requiring adjustment
must have all three adjustments: Zero, Range Gain + and
Range Gain .
(1) Set the rear panel CALIBRATION switch to ENABLE (if necessary).
(3) Ensure that nothing is connected to all INPUT binding posts.
29