TB 9-6625-134-24
(5) Press B.
(6) Press F3 to select PROBE B.
keys select Voltage, 1:1 and press ENTER.
(7) Using
(8) Press SCOPE.
(9) Press F1 to select READINGS ON.
(10) Press F2 READING.
(11) Press F1 to select READINGS 2.
keys select on B, ENTER, Vdc, and press ENTER.
(12) Using
(13) Press F2 READING.
(14) Press F1 to select READINGS 1.
keys select Off, and press ENTER.
(15) Using
(16) Press F4 CLOSE.
keys select Glitch: Off, Acquisition: Normal, Average: Off,
(18) Using
(19) Press MANUAL/AUTO to select MANUAL ranging (MANUAL in upper right
of screen).
(20) Press A, F1 to select INPUT A OFF.
(21) Press SAVE.
(22) Press F1 SAVE.
keys select Screen + Setup, and press ENTER.
(23) Using
(24) Press F1 EDIT NAME.
keys, F2 PREV # and F3 NEXT # set Save As: to CAL_B and
(25) Using
press F1 NAME OK.
keys select OK SAVE, and press ENTER.
(26) Using
8. Oscilloscope Vertical Accuracy
a. Performance Check
(1) Press TI keys as listed in (a) through (d) below.
SAVE.
(a)
F2 RECALL
(b)
(c)
Select CAL_A saved in 7. above using the
keys.
ENTER.
(d)
(2) Connect calibrator OUTPUT HI and LO to TI INPUT A and set output to
0 mVdc.
(3) Press TI keys as listed in (a) through (e) below.
(a) A, F4 INPUT A OPTIONS, using
20 kHz ENTER.
(b) CLEAR to exit softkey menu.
(c) s TIME ns, to select 10 mS/div.
(d) Using MOVE move the Input A ground level to the center grid line.