TB 9-6625-1356-24
Test instrument parameters
Performance specifications
Distortion
Models 651B and 652A (SG-763/U)
Range: 10 Hz to 2 MHz
Accuracy: <1%
Range: 2 to 5 MHz
Accuracy: <2%
Model 651A
Range: 5 to 10 MHz
Accuracy: <4%
Range: 10 Hz to 5 MHz
Accuracy: 1%
Approximately 2% at 10 MHz
Output Monitor
Range: 0.9 to 1 V (or 2.8 to 3.2 V)
Models 651B and 652A (SG-763/U) only
Accuracy: 2% FS
Output Monitor Flatness
Range: 3 and 1 V2
Models 651B and 652A (SG-763/U) only
Accuracy: 0.25%
Range: 0.3 to 0.3 mV
Accuracy: 0.75%
Range: 0.1 mV
Accuracy: 1.75%
Frequency Response
Range: 1 to 3 V3
Models 651B and 652A (SG-763/U)
Accuracy: 2%
Frequency: 20 Hz-4 MHz
Accuracy: 1%
Model 651A
Range: 1 to 3 V3
Frequency: 10 to 100 Hz
Accuracy: 3%
Frequency: 100 Hz to 4 MHz
Accuracy: 2%
Frequency: 4 to 10 MHz
Accuracy: 4%
1Procedure
limitation: Certification from -50 to +20 dB only.
2Amplitude
readjusted using expanded scale on output monitor.
3Specification is verified on the 3 V range only.