TB 9-6625-140-24
a. Performance Check
(1) Connect the function generator to TI V
Hz and COM terminals using 50
feedthrough termination. Set TI function switch to Hz.
(2) Set function generator for an 18 Hz output. Set the TI RANGE switch to the
first range in table 8. Adjust the function generator output until the meter gives a stable
reading.
(3) TI will indicate within the limits in the first line of table 8.
(4) Repeat steps (2) and (3) above for the remaining frequencies and RANGEs in
(5) Set function generator output to minimum and disconnect equipment setup.
Test Instrument
Applied
Limits
RANGE
(Hz)
(Hz)
Min
Max
18 Hz
20 Hz
17.989 Hz
18.011 Hz
180 Hz
200 Hz
179.89 Hz
180.11 Hz
1.8 kHz
2 kHz
1.7989 Hz
1.8011Hz
18 kHz
20 kHz
17.989 kHz
18.011 kHz
180 kHz
200 kHz
179.89 kHz
180.11 kHz
1.8 MHz
2 MHz
1.7989 MHz
1.8011 MHz
9 MHz
10 MHz
8.993 MHz
9.007 MHz
b. Adjustments. No adjustment can be made.
a. Performance Check
NOTE
If temperature and input adapter are not turned in with TI,
skip to paragraph 15 below and annotate DA label 163 with
"Temperature not calibrated" also.
(1) Connect the Temperature Probe to TI μm ATEMP and COM Input.
(2) Set TI controls as necessary to measure F temperature.
(3) Place TI Temperature Probe and Standard Thermometer near each other and
allow time for temperature stabilization.
(4) The TI indication must agree with the Thermometer within (3.0 F + 1 dgt).
(5) Set TI and Standard Thermometer controls as necessary to measure C
temperature.
(6) The TI indication must agree with the Thermometer within (1.5 C + 1 dgt).