Table 8. Frequency Accuracy

Calibration Procedure For Digital Multimeter Model Eedm586C Page Navigation
  4    5    6    7    8  9
TB 9-6625-140-24
15. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label 163 annotated block (a), with parameter, RPM IP and
RPM IG and block (b), not calibrated in accordance with TB 750-25. If temperature and
input adapter were not turned in with TI, annotate DA label 163 with "Temperature not
calibrated" also.
10