TB 9-6625-149-50
8. Calibrator Accuracy
a. Performance Check
(1) Set TI POWER switch to OFF and remove V875 (fig. 1) from TI.
(2) Set TI POWER switch to ON.
(3) Connect ac/dc voltmeter (A2) between TI CAL OUT connector and chassis
ground, using leads supplied with ac/dc voltmeter and, if necessary, adapter (B2).
(4) Turn TI AMPLITUDE CALIBRATOR switch to settings listed in table 4. If
ac/dc voltmeter does not indicate within limits specified, perform b below.
(5) Set TI POWER switch to OFF and replace V875.
(6) Set TI POWER switch to ON.
7