TB 9-6625-152-24
Table 2. Minimum Specifications of Equipment Required.
Manufacturer and model
Common name
Minimum use specifications
(part number)
Fluke, Model PM6681/656
Accuracy:
1 count LSD
(PM6681/656)
CALIBRATOR
Ac volts:
Fluke, Model 5720A
Range:
30 mV to 30 V rms
(5700A/EP) (p/o MIS-35947);
20 Hz to 50 kHz
w/ amplifier, Fluke, Model
Accuracy: 0.75 %
5725A/AR (5725A/AR)
Dc volts:
Range:
0 to 40 V
Accuracy: 0.25%
Range:
-60 to +13.01 dBm
Agilent, Model 33250A
20 Hz to 50 kHz
(33250A)
Accuracy: 0.625 dB
MEASURING RECEIVER
Measuring receiver system
Flatness: 500 kHz to 999 MHz
N5530S consisting of: Spectrum
Analyzer Agilent, Model
Power:
0 to +30 dB 0.103 dB
E4440A (E4440A), Power Meter
0 to -110 0.375 dB
Agilent, Model E4419B
-110 to -127 0.625 dB
(E4419B), and Sensor Module
Accuracy: 0.5 dB
Agilent Model N5532A opt. 504
FM:
(504)
Range:
10 to 90 kHz
Accuracy: 1% <15 kHz mod rate
1.75 >15 kHz mod rate
AM:
Range:
0.1% 90%
Accuracy: 1.25
Range:
1 to 999.999 MHz
Noise Floor: <50 dB
Audio analysis:
Range:
50 Hz to 15 kHz
Accuracy: 1 dB
Ac volts:
Hewlett-Packard, Model 3458A
Range:
20 Hz to 25 kHz
(3458A)
1 mV to 2.5 V rms
Accuracy: 25 mV or .75%
<10 kHz
1.25% >10 to 25 kHz
Dc volts:
Range:
4.096 to 35 V dc
Accuracy: (.0006 %) .005%
Range:
50 Hz
Agilent, Model 54831M
Accuracy: <125 ns risetime
(OS-303/G)
POWER SPLITTER
Insertion loss: -6dB
Weinschel, Model 1870A
Output tracking between ports:
(7916839)
10 MHz to 1 GHz: 0.15 dB
5