TB 9-6625-154-40
(5) Using the TI arrow keys adjust the TI O/P Amplitude until the power meter
display matches the first Meter Indication listed in table 10 below and press TI Softkey,
ACCEPT CALIB.
(6) Repeat technique of (3)(m) and (5) above for remaining rows in table 10.
Table 10. Leveled Sine HF Gain Adjust 250 MHz.
Power Meter
Cal Point
Frequency (MHz)
TI level (V p-p)
Indication
Target 1
250
1.0000
3.979
1.5000
Target 2
250
7.501
2.0000
Target 3
250
10.000
3.0000
Target 4
250
13.522
5.5000
Target 5
250
18.787
(7) Set TI as listed in (a) and (b) below:
(a) Softkey, FLAT.
(b) Softkey, TARGET 1.
(9) Using the TI arrow keys adjust the TI O/P Amplitude until the power meter
display matches the first Meter Indication listed in table 11 below and press TI Softkey,
ACCEPT CALIB.
(10) Repeat technique of (7)(b) and (9) above for remaining rows in table 11.
(11) Press TI Softkey, NEXT FREQ and repeat technique of (7)(b) through (10) above
for 10 MHz.
(12) Repeat technique of (11) above for 50, 250, 550, 800 MHz and 1.0 GHz.
Table 11. Leveled Sine HF Gain Adjust 50 kHz to 1 GHz.
TI level
Power Meter
Cal Point
(V p-p)
Indication
Target 1
3.0000
13.521
1.0072
Target 2
4.042
316.49 m
Target 3
-6.013
316.49 m
Target 4
-6.013
(13) Set TI as listed in (a) through (e) below:
(a)
Softkey, EXIT.
(b)
Softkey, SINE.
(c)
Softkey, 3 GHz Sine.
(d)
Softkey, TARGET 1.
(e)
OUTPUT ON.
(15) Using the TI arrow keys adjust the TI O/P Amplitude until the power meter
display matches the first Meter Indication listed in table 12 below and press TI Softkey,
ACCEPT CALIB.
(16) Repeat technique of (13)(d) and (15) above for remaining rows in table 12.