TB 9-6625-156-24
o. Click RUN BIT TEST on CORRECTION CONTROL tab and verify "Pass" condition
for all tests in the CONNECTION LOG. A new TI or a recently repaired TI may fail "RF
PATH" and "CALIBRATION" because a default calibration file is being used.
p. Select the RF CALIBRATION tab and wait for "Warmup Delay" or "Temperature
Not Stable" under the BOARD TEMPERATURE bar to change to read "Temperature
Stable". This may initially take 20 minutes.
q. Click GET FROM UUT. Serial number, coefficients and other parameters are read
from TI. TI is now prepared for calibration.
NOTE
The TI serial number may be changed later in paragraph 14.
6