TB 9-6625-157-40
(3) Calculate the resistance corresponding to ĆAvg. Temp using formula below:
(4) Record the current value of the RZ (current resistance offset) from the TI
(5) Calculate the new RZ by algebraically adding the ĆResistance to the current RZ
using formula below:
(6) Adjust RZ by pressing TI keys as listed in (a) through (f) below. (R)
[ESCAPE]
(a)
<2cal>
(b)
Using key pad, input new RZ.
(c)
[ENTER]
(d)
[ENTER]
(e)
<2yes>
(f)
(8) Remove thermometer and PRT insertion plug from TI.
(9) Replace the piston cylinder module, mass set and automatic mass handler.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.