TB 9-6625-1873-50
7. VSWR
a. Performance Check
Figure 1. VSWR - equipment setup.
(2) Adjust signal generator and square-wave generator (A1 and A2) for 200 MHz
output, square-wave modulated at 1 kHz. Standardize equipment for VSWR measurement.
(3) Measure VSWR. VSWR will not exceed 1.2.
(4) Repeat technique of (1) through (3) above for model TP-93.
b. Adjustments. No adjustments can be made.
4