TB 9-6625-1932-24
APPENDIX A
TEST INSTRUMENT SPECIFICATIONS
Power Sensors - Continued
Standard limitations
from test report
indications
VSWR
(Kb factor)
max
Model
Power range
Hewlett-Packard 10 MHz to 18 GHz 100 pW to 10 W 1.0 GHz: 1.9%
8481D1
10 to 30 MHz: 1.40
2.0 to 6.0 GHz: 1.8%
30 MHz to 4 GHz: 1.15
8.0 GHz: 2.0%
4 to 10 GHz: 1.20
10.0 GHz: 2.2%
10 to 15 GHz: 1.30
12.0 GHz: 2.8%
15 to 18 GHz: 1.35
14.0 GHz: 3.2%
16.0 GHz: 3.4%
18.0 GHz: 3.7%
Hewlett-Packard 10 MHz to 18 GHz 100 W to 3 W
3% below 25 dBm
10 MHz to 8 GHz: 1.20
8481H
5%, 25 to 35 dBm
8 to 12.4 GHz: 1.25
12.4 to 18 GHz: 1.30
0.3 W to 100 mW 0.1 MHz to 4 GHz: 100 to 300 kHz: 1.60
Hewlett-Packard 100 kHz to 4.2 GHz2
8482A
3%
300 kHz to 1 MHz: 1.20
1 MHz to 2 GHz: 1.10
2 to 4.2 GHz: 1.3
0.1 to 30 MHz: 2.8% 0.1 to 2 GHz: 1.10
8482B
Hewlett-Packard 100 kHz to 4.2 GHz
1 mW to 25 W
50 MHz: 2.7%
2 to 4.2 GHz: 1.18
100 to 1000 MHz::
3.3%
1000 to 4000 MHz::
3.1%
100 W to 3 W
3% below 25 dBm
Hewlett-Packard 100 kHz to 4.2 GHz2
100 kHz to 4.2 GHz: 1.20
8482H
5%, +25 to +35 dBm
0.3 W to 100 mW
8483A
Hewlett-Packard 100 kHz to 2 GHz
0.1 MHz to 2 GHz: 100 to 600 kHz: 1.80
(75 )2
3%
600 kHz to 2 GHz: 1.18
0.1 nW to 10 W 10 MHz to 12.4 GHz: 10 to 30 MHz: 1.4
8484A3
Hewlett-Packard 10 MHz to 18 GHz
3%
30 MHz to 4 GHz: 1.15
4 to 10 GHz: 1.2
12.4 to 14 GHz:
3.24%
10 to 18 GHz: 1.30
14 to 16 GHz:
3.4%
16 to 18 GHz:
3.30%
See footnotes at end of table.
A-3