TB 9-6625-194-24
Table 6. Model i-400s Ac Current.
Calibrator
TI Applied
Range
applied
calibrator range
current
Min
Max
(A)
(A)
(A ac)
(V ac)
(mV ac)
(mV ac)
1.2 @ 400 Hz
20
12
117.45
122.55
2
1.8 @ 45 Hz
2
27
264.45
275.55
1.8 @ 400 Hz
2
27
264.45
275.55
0.24 @ 45 Hz
20
36
352.65
367.35
0.24 @ 400 Hz
20
36
352.65
367.35
400
0.8 @ 45 Hz
20
120
117.56
122.44
0.8 @ 400 Hz
20
120
117.56
122.44
1.86@ 45 Hz
20
279
273.38
284.62
1.86 @ 400 Hz
20
279
273.38
284.62
0.24@ 45 Hz
100
360
352.76
367.24
0.24 @ 400 Hz
100
360
352.76
367.24
Connect the single turn coil to the Ammeter Calibrator.
1
2 Connect the 15-turn coil to the Ammeter Calibrator.
Table 7. Model 80i-500s Ac Current.
Calibrator
TI Applied
Range
applied
calibrator range
current
Min
Max
(A)
(V ac)
(A)
(A ac)
(V ac)
(V ac)
20 1
1.9 @ 10 Hz
285
0.26505
0.30495
0.26 @ 15 Hz
100
390
0.32370
0.45630
0.326 @ 20 Hz
20
489
0.45477
0.52323
0.316 @ 40 Hz
20
474
0.44082
0.50718
500
1.2 @ 60 Hz
2
18
16.8 m
19.2 m
0.6 @ 60 Hz
20
90
85.500 m
94.500 m
0.326 @ 60 Hz
100
489
0.47922
0.49878
0.3 @ 3 kHz
20
45
40.100 m
49.900 m
0.326 @ 400 Hz
20
489
0.45477
0.52323
Connect the 15-turn coil to the Ammeter Calibrator.
1
Table 8. Model 80i-600 and Model 80i-600A Ac Current.
Calibrator
TI Applied
Range
applied
calibrator range
current
Min
Max
(A)
(V ac)
(A)
(A ac)
(A ac)
(A ac)
1
600
0.36 @ 40 Hz
100
540
0.5238
0.5562
0.28 @ 60 Hz
100
420
0.4116
0.4284
0.12 @ 60 Hz
100
180
0.1764
0.1836
0.36 @ 400 Hz
100
549
0.53802
0.55998
0.33 @ 1 kHz
20
49.5
49.51 m
50.49 m
0.16 @ 5 kHz
20
24
23.28 m
24.72 m
Connect the 15-turn coil to the Ammeter Calibrator.
1
6