TB 9-6625-1957-35
(4) Repeat technique of (2) (b) and (3) above, using outputs and settings listed in
Test instrument
RANGE/FUNCTION
Calibrator
Indication limits
output
setting
Min
Max
190 A
200 A
189.59
190.41
1.9 mA
2 mA
1.8959
1.9041
19 mA
20 mA
18.959
19.041
190 mA
200 mA
189.59
190.41
1.9
2000 mA
1895.9
1904.1
(5) Reduce calibrator output to minimum.
b. Adjustments. No other adjustments can be made.
a. Performance Check
(1) Connect calibrator OUTPUT HI to TI VOLTS/Ω input and calibrator OUTPUT
LO to TI LOW input.
(2) Press TI pushbuttons as listed in (a) through (c) below:
(a) kΩ (in).
(b) 20 kΩ range (in).
(c) HI LO (out).
(3) Set calibrator for a 19k Ω nominal output.
(4) Rotate calibrator knob below EDIT FIELD pushbutton to adjust calibrator
display indication to equal TI indication. If calibrator err display does not indicate within
limits specified in first row of table 11, perform b below.
(5) Repeat technique of (2) (b) and (c), (3) and (4) above, using calibrator outputs
and TI settings and indications listed in table 11. Calibrator err display will indicate
Test instrument
Calibrator
err display
HI-LO
Nominal
indications
Range
pushbuttons
output
(%)
pushbuttons
LO
.166
kΩ
kΩ
201
19
HI
.166
kΩ
kΩ
20
19
LO
.166
Ω
Ω
200
190
LO
.166
kΩ
kΩ
2
1.9
HI
.166
kΩ
kΩ
2
1.9
HI
.166
kΩ
kΩ
2002
190
LO
.166
kΩ
kΩ
200
190
LO
.316
kΩ
MΩ
2000
1.9
HI
.166
kΩ
MΩ
2000
1.9
HI
.516
MΩ
MΩ
20
19
Set calibrator 2 WIRE COMP to on.
1
Set calibrator 2 WIRE COMP to off.
2
14 CHANGE 3