TB 9-6625-1961-24
Adjustments required to calibrate the TI are included in this procedure.
Additional
maintenance information is contained in the manufacturer's manual.
d. When indications specified in paragraphs 8 through 14 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 14. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Set TI DC
SWITCH (internal (fig. 1)) to DC (up) position.
b. Install TI into power module.
c. Connect power module to autotransformer, and leave connected throughout entire
d. Connect autotransformer to a 115 V ac source and adjust autotransformer for a 115 V output.
e. Energize power module by pulling the power button out (located on left edge of front panel).
f. Set TI model switch to STD AMPL and turn AMPLITUDE switch to 10. Allow at
least 20 minutes for equipment to warm-up.
g. Depress VARIABLE (out) control. DEFLECTION ERROR display will be extinguished.