TB 9-6625-1965-24
and RANGE 200
(2) Press TI FUNCTION K
pushbuttons.
nominal output.
(3) Set calibrator for a 190
(4) Rotate calibrator knob below EDIT FIELD pushbutton to adjust calibrator
display indication to equal TI indication. If calibrator err display does not indicate within
limits specified in first row of table 6, perform b below.
(5) Repeat technique of (2) through (4) above, using calibrator outputs and TI
indications listed in table 6. If calibrator err display does not indicate within limits
Test instrument
Calibrator
err
RANGE
display indication
(%)
pushbutton settings
Nominal output
190
0.253
200
1.9 k
0.253
2 k
19 k
0.253
20 k
190 k
0.253
200 k
1900 k
0.253
2000 k
19 M
0.553
20 M
b. Adjustments
(1) Set calibrator for 19.00 M . Adjust R55 20M (fig. 1) until TI indicates actual
(2) Adjust calibrator for 19.00 k . Adjust R12 KOHM (fig. 1) until TI indicates
12. Resistance (Model 8000A/MAS Option Only)
a. Performance Check
(1) Connect TI V-
input to calibrator OUTPUT HI and TI COMMON input to
calibrator OUTPUT LO.
(2) Press TI FUNCTION K
and RANGE 200
pushbuttons.
nominal output.
(3) Set calibrator for a 190
(4) Rotate calibrator knob below EDIT FIELD pushbutton to adjust calibrator
display indication to equal TI indication. If calibrator err display does not indicate within
limits specified in first row of table 7, perform b below.
(5) Repeat technique of (2) through (4) above, using calibrator outputs and TI
indications listed in table 7. If calibrator err display does not indicate within limits