TB 9-6625-1965-24
(10) Momentarily set calibrator to OPERATE. If digital multimeter does not
indicate less than 1 V dc with +22 mA applied to TI, adjust R7 +COMP (fig. 2) until digital
(11) Set mASEC mA switch to mA.
(12) Set calibrator for an initial 17 mA output and, if necessary, rotate calibrator
knob below EDIT FIELD pushbutton to adjust calibrator for a TI indication of 17 mA.
(13) Set calibrator to STANDBY and set mASEC mA switch to mASEC.
(14) Momentarily set calibrator to OPERATE. If digital multimeter does not
indicate between -13 and -17 V dc, adjust R7 +COMP (fig. 2) until digital multimeter does
indicate between -13 and -17 V dc.
(15) Set calibrator to STANDBY, reverse leads on calibrator and repeat (5) through
(14) above. If digital multimeter does not indicate less than 1 V dc when -22 mA is
applied and between -13 and -17 V dc when -17 mA is applied to TI, adjust R6 -COMP
(fig. 2) until both indications are obtained.
11