TB 9-6625-1967-35
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
a. Performance Check
(1) Connect multimeter common lead to TI TP6332 (GND) and positive lead to
TP6256 (fig. 5). If multimeter
does not
indicate between 24.75 and
25.25 V
dc,
perform b below.
NOTE
Leave common lead connected to TP6332 (GND) throughout
this performance check.
(2) Repeat technique of (1) above for test points listed in table 6. If multimeter does
not indicate within limits specified, perform adjustment listed in table 6.
Test instrument
Adjustments
(V dc)
test points
Min
Max
TP6411 (+10)
+9.9
+10.1
R6514 (+10ADJ) (R)
TP6227 (+5)
+4.75
+5.25
---
TP6218 (-5)
-4.75
-5.25
---
P66 (PIN 6) +100V
+90
+110
---
P66 (PIN 7) +165V
+155
+175
---
b. Adjustments.
Adjust R6358 (fig. 5) for a 25 V dc
indication on
multimeter (R).
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.