TB 9-6625-1979-24
Paragraph
Page
Equipment setup ..................................................
FET bias, output zero, and output bias ..............
Low and high frequency gain ..............................
Compression .........................................................
Frequency compensation and risetime ...............
Attenuator compensation ....................................
Final procedure ....................................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Power Supply and Probe, Tektronix, Types 1101 and P6201.
The
manufacturers' manuals were used as the prime data source in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Power Supply, Tektronix, Type 1101
Output voltages:
+15 V 0.75%
Range and accuracy
-15 V 1.5%
+5 V 2.0%
Probe, Tektronix, Type P6201
Risetime (probe only)
390 ps or less