TB 9-6625-1980-24
(8) Press POWER OPR pushbutton.
(9) If multimeter does not indicate between 99.98800 and 100.01200, perform b (31)
and (32) below.
(10) Press POWER STDBY/RESET pushbutton.
(11) Press MODE 100-1000 pushbutton.
(12) Set TI decade dials to indicate 700.0000.
(13) Press POWER OPR pushbutton.
(14) If multimeter does not indicate between 699.916 and 700.084, perform b (28)
and (29) below.
(15) Press POWER STDBY/RESET pushbutton.
b. Adjustments
(1) Press POWER STDBY/RESET pushbutton.
(2) Disconnect TI from multimeter.
WARNING
Use only insulated adjusting tools when making adjustments.
Do not let test probe tips touch the protective cover.
(4) Press POWER OPR pushbutton.
(5) If multimeter does not indicate 0 10 V, adjust A4A1R10 REF ADJ (fig. 2) until
multimeter indicates 0 (R).
(6) Press POWER STDBY/RESET pushbutton.
(7) Disconnect leads connected in (3) above and reconnect TI OUTPUT
terminals to multimeter.
(8) Repeat a (2) and (3) above. If TI does not indicate within limits, perform (10)
and (11) below.
(9) Press POWER STDBY/RESET pushbutton.
(10) Press MODE 10-100 and POWER OPR pushbuttons.
(11) Perform ladder No. 1 adjustments, settings listed in table 4. For each setting,
adjust respective adjustment for a nominal indication on multimeter within 10 V. Repeat
a (2) through (5) above. If TI decade dials do not indicate within limits, perform b (12)
through (25) below.
8