TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Amprobe Instruments, Model 15XP-A - Continued
Table B-16. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
190 μ
200 μ
187.9 μ
192.1 μ
2000 μ
μ
μ
1.9 m
1879
1921
19 m
20 m
18.79 m
19.21 m
190 m
200 m
187.9 m
192.1 m
1.9
2
1.859
1.941
Table B-17. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
190 μ @ 45 Hz
200 μ
186.6 μ
193.4 μ
190 μ @ 500 Hz
200 μ
186.6 μ
193.4 μ
1900 μ @ 45 Hz
2000 μ
μ
μ
1866
1934
1900 μ @ 500 Hz
2000 μ
μ
μ
1866
1934
19 m @ 45 Hz
20 m
18.66 m
19.34 m
19 m @ 500 Hz
20 m
18.66 m
19.34 m
190 m @ 45 Hz
200 m
186.6 m
193.4 m
190 m @ 500 Hz
200 m
186.6 m
193.4 m
1.9 @ 45 Hz
2
1.847
1.953
1.9 @ 500 Hz
2
1.847
1.953
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
200
-1.211% (187.7)
+1.211% (192.3)
1.9 k
2k
-1.211% (1.877 k)
+1.211% (1.923 k)
19 k
20 k
-1.211% (18.77 k)
+1.211% (19.23 k)
190 k
200 k
-1.211% (187.7 k)
+1.211% (192.3 k)
1.9 M
2M
-1.211% (1.877 M)
+1.211% (1.923 M)
19 M
20 M
-3.263% (18.38 M)
+3.263% (19.62 M)
1
+25.00% (125 M)1
100 M
2000 M
-25.00% (75 M)
Subtract 10 MΩ from TI displayed value. Result must be within specified limits.
1
B-6