TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 16XL - Continued
Table B-21. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
180 m @ 50 Hz
200 m
177.8 m
182.2 m
180 m @ 500 Hz
200 m
177.8 m
182.2 m
1.8 @ 50 Hz
2
1.778
1.822
1.8 @ 500 Hz
2
1.778
1.822
2 @ 400 Hz
20
1.94
2.06
6 @ 400 Hz
20
5.90
6.10
10 @ 400 Hz
20
9.86
10.14
14 @ 400 Hz
20
13.82
14.18
18 @ 50 Hz
20
17.78
18.22
18 @ 500 Hz
20
17.78
18.22
180 @ 50 Hz
200
177.8
182.2
180 @ 500 Hz
200
177.8
182.2
540 @ 50 Hz
600
528
552
540 @ 50 Hz
600
528
552
Table B-22. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
18 m
20 m
17.81 m
18.19 m
180 m
200 m
178.1 m
181.9 m
101
9
8.79
9.21
Move lead from TI mA terminal to TI 10 A terminal.
1
Table B-23. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
18 m @ 50 Hz
20 m
17.60 m
18.40 m
18 m @ 500 Hz
20 m
17.60 m
18.40 m
180 m @ 50 Hz
200 m
176.0 m
184.0 m
180 m @ 500 Hz
200 m
176.0 m
184.0 m
9 @ 50 Hz
10
8.73
9.27
9 @ 500 Hz
10
8.73
9.27
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
200
-1.1579% (187.8)
+1.1579% (192.2)
1.9 k
2k
-1.0526%(1.880 k)
+1.0526% (1.920 k)
19 k
20 k
-1.0526%(18.80 k)
+1.0526% (19.20 k)
190 k
200 k
-1.0526%(188.0 k)
+1.0526% (192.0 k)
1.9 M
2M
-1.0526%(1.880 M)
+1.0526% (1.920 M)
19 M
20 M
-3.0526% (18.42 M)
+3.0526% (19.58 M)
B-8