TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 2015
Test instrument parameters
Performance specifications
DC VOLTAGE
Range: 0 to 1000 V ac
Accuracy: (0.25% rdg + 2 dgts)
AC VOLTAGE
Range: 0 to 750 A ac
Accuracy: 50 to 100 Hz: (1% of rdg + 5 dgts), 400 mV ac rng;
50 to 500 Hz: (0.8% of rdg + 3 dgts), 4 to 750 V ac rngs;
500 Hz to 1 kHz: (1.2% of rdg + 5 dgts), 4 VAC rng;
(1% of rdg + 5 dgts), 40 and 400 V ac rngs;
(1.2% of rdg + 5 dgts), 750 V ac rng
DC CURRENT
Range: 0 to 20 A dc
Accuracy: (0.8% of rdg + 2 dgts), 4 to 400 mA dc rngs;
(1.2% of rdg + 4 dgts), 20 A dc rng
AC CURRENT
Range: 0 to 20 A ac, 50 Hz to 1 kHz
Accuracy: (1.2% of rdg + 4 dgts), 4 to 400 mA ac rngs;
(1.5% of rdg + 5 dgts), 20 A ac rng
Range: 0 to 40 MΩ
Accuracy: (0.8% of rdg + 4 dgts), 400 Ω rng;
(0.7% of rdg + 2 dgts), 4 to 400 kΩ rngs;
(1% of rdg + 4 dgts), 4 MΩ rng;
(1.5% of rdg + 4 dgts), 40 MΩ rng
Range: 0 to 40 μF
Accuracy: (1% of rdg + 40 dgts), 4 nF rng;
(1% of rdg + 4 dgts), 40 nF to 4 μF rngs;
(1% of rdg + 4 dgts), <20 μF on the 40 μF rng;
(5% of rdg + 8 dgts), >20 μF on the 40 μF rng
Range: 0 to 400 kHz
Sensitivity: 40 mV, 100 Hz to 100 kHz; 400 mV, 100 to 400 kHz
Accuracy: (0.1% of rdg + 10 dgts), 100 Hz rng;
(0.1% of rdg + 4 dgts), 1 and 10 kHz rngs;
(0.1% of rdg + 8 dgts), 100 kHz rng;
(0.1% of rdg + 20 dgts), 400 kHz rng
Table B-28. DC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V dc)
(V dc)
(V dc)
Min
Max
360 m
400 m
358.9 m
361.1 m
3.6
4
3.589
3.611
12
40
11.95
12.05
28
40
27.91
28.09
36
40
35.89
36.11
-36
40
-36.11
-35.89
360
400
358.9
361.1
900
1000
896
904
B-10