TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model 2020 - Continued
Table B-37. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
90 m @ 50 Hz
100 m
88.12 m
91.88 m
90 m @ 400 Hz
100 m
88.12 m
91.88 m
900 m @ 50 Hz
1000 m
881.2 m
918.8 m
900 m @ 400 Hz
1000 m
881.2 m
918.8 m
900 m @ 2 kHz
1000 m
881.2 m
918.8 m
9 @ 50 Hz
10
8.812
9.188
9 @ 400 Hz
10
8.812
9.188
9 @ 2 kHz
10
8.812
9.188
90 @ 50 Hz
100
88.12
91.88
90 @ 400 Hz
100
88.12
91.88
90 @ 2 kHz
100
88.12
91.88
900 @ 50 Hz
1000
881.2
918.8
900 @ 400 Hz
1000
881.2
918.8
900 @ 2 kHz
1000
881.2
918.8
Table B-38. DC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A dc)
(A dc)
Min
Max
900 μ
1000 μ
890.8 μ
909.2 μ
9m
10 m
8.954 m
9.046 m
90 m
100 m
89.54 m
90.46 m
1000 m1
900 m
890.8 m
909.2 m
9
10
8.909
9.091
10
20
9.899
10.101
Move lead from TI 100 mA terminal to TI 10 A terminal.
1
Table B-39. AC Current.
Calibrator output
Test instrument
Applied
Range
Limits (A ac)
(A ac)
(A ac)
Min
Max
900 μ @ 50 Hz
1000 μ
883.9 μ
916.1 μ
900 μ @ 1 kHz
1000 μ
883.9 μ
916.1 μ
9 m @ 50 Hz
10 m
8.839 m
9.161 m
9 m @ 1 kHz
10 m
8.839 m
9.161 m
90 m @ 50 Hz
100 m
88.42 m
91.58 m
90 m @ 1 kHz
100 m
88.42 m
91.58 m
1
900 m @ 50 Hz
1000 m
884.2 m
915.8 m
900 m @ 1 kHz
1000 m
884.2 m
915.8 m
9 @ 50 Hz
10
8.844
9.156
9 @ 1 kHz
10
8.844
9.156
10 @ 50 Hz
20
9.827
10.173
10 @ 1kHz
20
9.827
10.173
Move lead from TI 100 mA terminal to TI 10 A terminal.
1
B-14