TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Amprobe Instruments, Model 38XR-A - Continued
Table B-96. DC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V dc)
(V dc)
(V dc)
Min
Max
900 m
1000 m
897.2 m
902.8 m
3
10
2.987
3.013
-3
10
-3.013
-2.987
5
10
4.982
5.018
-5
10
-5.018
-4.982
7
10
6.977
7.023
-7
10
-7.023
-6.977
9
10
8.972
9.028
-9
10
-9.028
-8.972
90
100
89.72
90.28
900
1000
897.2
902.8
Table B-97. AC, AC+DC Voltage and dBm.
Calibrator output
Test instrument
Applied
Range
Limits
(V ac)
(V ac)
Min
Max
900 m @ 45 Hz
1000 m
888.2 mV
911.8 mV
900 m @ 2 kHz
1000 m
881.0 mV
919.0 mV
3 @ 45 Hz
10
2.954 V
3.046 V
3 @ 2 kHz
10
2.930 V
3.070 V
5 @ 45 Hz
10
4.930 V
5.070 V
5 @ 2 kHz
10
4.890 V
5.110 V
7 @ 45 Hz
10
6.906 V
7.094 V
7 @ 2 kHz
10
6.850 V
7.150 V
9 @ 45 Hz
10
8.882 V
9.118 V
9 @ 2 kHz
10
8.810 V
9.190 V
90 @ 45 Hz
100
88.82 V
91.18 V
90 @ 2 kHz
100
88.10 V
91.90 V
90 @ 500 Hz1
67.3 V2
112.7 V2
100
675 @ 45 Hz
750
660.5 V
689.5 V
675 @ 1 kHz
750
660.5 V
689.5 V
675 @ 500 Hz1
634.7 V2
715.3 V2
750
-13 dBm @ 45 Hz3
dBm
-13.17 dBm
-12.83 dBm
-13 dBm @ 4 kHz
dBm
-13.17 dBm
-12.83 dBm
-13 dBm @ 6 kHz
dBm
-13.41 dBm
-12.59 dBm
-13 dBm @ 6 kHz
dBm
-13.41 dBm
-12.59 dBm
+50 dBm @ 45 Hz
dBm
49.57 dBm
50.43 dBm
+50 dBm @ 4 kHz
dBm
49.57 dBm
50.43 dBm
+50 dBm @ 6 kHz
dBm
48.67 dBm
51.33 dBm
+50 dBm @ 10 kHz
dBm
48.67 dBm
51.33 dBm
900 m @ 45 Hz4
1000 m
885.5 m
914.5 m
900 m @ 2 kHz
1000 m
876.5 m
923.5 m
See footnotes at end of table.
B-40