TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
Wavetek, Model HD160
Test instrument parameters
Performance specifications
DC VOLTAGE
Range: 0 to 1000 V dc
Accuracy: 1 V dc rng, (0.25% rdg + 3 dgts);
10 to 1000 V dc rngs, (0.25% rdg + 1 dgt)
AC VOLTAGE
Range: 0 to 1000 V ac
Accuracy: 10 V ac rng, @ 45 Hz to 10 kHz, (1.5% rdg + 5 dgts);
100 V ac rng, 2 to 29.9 V ac @ 45 Hz to 10 kHz, (1.5% rdg + 1 dgt);
2 to 29.9 V ac @ 10 kHz to 30 kHz, (2.5% rdg + 5 dgts);
30 to 100 V ac @ 45 Hz to 1 kHz, (1.5% rdg + 5 dgts);
1000 V ac rng, 45 Hz to 1 kHz, (1.5% rdg + 1 dgt)
DC CURRENT
Range: 0 to 20 A dc
Accuracy: 10 mA dc to 10 A dc rngs, (0.75% rdg + 5 dgts);
20 A dc rng, (0.75% rdg + 3 dgts)
AC CURRENT
Range: 0 to 20 A ac
Accuracy: 10 mA ac to 10 A ac rngs, (1.75% rdg + 5 dgts);
20 A ac rng, (1.75% rdg + 3 dgts)
Range: 0 to 40 MΩ
Accuracy: 1 kΩ to 1 MΩ rngs, (0.5% rdg + 1 dgt);
10 MΩ rng, (0.5% rdg + 2 dgts);
40 MΩ rng, (1.5% rdg + 2 dgts)
Table B-190. DC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V dc)
(V dc)
(V dc)
Min
Max
900 m
1
897.4 m
902.6 m
3
10
2.991
3.009
5
10
4.986
5.014
7
10
6.981
7.019
9
10
8.976
9.024
-9
10
-9.024
-8.976
90
100
89.76
90.24
900
1000
897.6
902.4
Table B-191. AC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V ac)
(V ac)
(V ac)
Min
Max
9 @ 50 Hz
10
8.860
9.140
9 @ 400 Hz
10
8.860
9.140
9 @ 1 kHz
10
8.860
9.140
9 @ 10 kHz
10
8.860
9.140
90 @ 50 Hz
100
88.60
91.40
90 @ 1 kHz
100
88.60
91.40
27 @ 50 Hz
100
26.58
27.42
27 @ 10 kHz
100
26.58
27.42
27 @ 11 kHz
100
26.27
27.73
27 @ 30 kHz
100
26.27
27.73
900 @ 50 Hz
1000
886.4
913.6
900 @ 1 kHz
1000
886.4
913.6
B-75