TB 9-6625-199-24
APPENDIX B
PERFORMANCE TABLES
(All Manufacturers), Model M2625 - Continued
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
400
-1.263% (187.6)
+1.263% (192.4)
1.9 k
4k
-1.623% (1.876 k)
+1.623% (1.924 k)
19 k
40 k
-1.623% (18.76 k)
+1.623% (19.24 k)
190 k
400 k
-1.623% (187.6 k)
+1.623% (192.4 k)
1.9 M
4M
-1.623% (1.876 M)
+1.623% (1.924 M)
19 M
40 M
-2.263% (18.57 M)
+2.263% (19.43 M)
Test instrument
Applied
Range
Limits (F)
(F)
(F)
Min
Max
10 n
40 n
9.55 n
10.45 n
100 n
400 n
96.5 n
103.5 n
1μ
4μ
0.965 μ
1.035 μ
10 μ
40 μ
9.65 μ
10.35 μ
90 μ
100 μ
86.3 μ
93.7 μ
Table B-202. Temperature.
Cool/Heat source
Test instrument
Limits
Applied
Range
Min
Max
0 F
F
-5 F
5 F
650 F
F
635 F
665 F
Test instrument
Applied
Range
Limits (Hz)
(Hz)
(Hz)
Min
Max
700 mV @ 100 Hz
Auto
99.4
100.6
700 mV @ 1 kHz
Auto
9.94 k
10.06 k
700 mV @ 100 kHz
Auto
99.4 k
100.6 k
700 mV @ 1 MHz
Auto
0.994 M
1.006 M
1.5 V @ 10 MHz
Auto
9.94 M
10.06 M
B-79