TB 9-6625-1995-24
10th marker with 10th vertical graticule line. If oscilloscope calibrator err display readout
does not indicate within 4%, perform b below.
(11) Set TIME/DlV OR DLY TIME and DLY'D time/division switches to 5 ns
PUSH FOR DLY'D.
(12) Set oscilloscope calibrator MARKER output for 5 nS/D.
(13) Adjust POSITION control to align 2d time marker with 2d vertical graticule
line. Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to align
10th marker with 10th vertical graticule line. If oscilloscope calibrator err display readout
does not indicate within 3%, perform b below.
(14) Set TIME/DIV OR DLY TIME and DLY'D time/division switches to 10 ns
PULL FOR ALT.
(15) Set oscilloscope calibrator MARKER output for 10 nS/D. Adjust vertical
amplifier POSITION control to display both the delaying and delayed sweeps on
oscilloscope.
(16) Adjust POSITION control to align 2d time marker with 2d vertical graticule
line. Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to align
10th marker with 10th vertical graticule line. If oscilloscope calibrator err display readout
does not indicate within 3%, perform b below.
(17) Repeat technique of (14) through (16) above for settings listed in table 4. If
oscilloscope calibrator err display readout does not indicate within 2%, perform b below.
Test instrument
TIME/DIV OR DLY TIME
and DLY'D time/division
PULL FOR ALT
Oscilloscope calibrator marker
switch settings
output settings
20
ns
20
nS/D
50
ns
50
nS/D
s
S/D
0.1
0.1
s
S/D
0.2
0.2
s
S/D
0.5
0.5
s
S/D
1
1
s
S/D
2
2
s
S/D
5
5
s
S/D
10
10
s
S/D
20
20
s
S/D
50
50
0.1
ms
0.1
mS/D
0.2
ms
0.2
mS/D
0.5
ms
0.5
mS/D
1
ms
1
mS/D
2
ms
2
mS/D
5
ms
5
mS/D
10
ms
10
mS/D
20
ms
20
mS/D
50
ms
50
mS/D
0.1
s
0.1
S/D
0.2
s
0.2
S/D