TB 9-6625-1995-35
Test instrument
DELAY TIME
DELAY TIME MULT
MULT
dial indications
dial settings
Min
Max
7.0
6.93
7.07
6.0
5.92
6.08
5.0
4.91
5.09
4.0
3.90
4.10
3.0
2.89
3.11
2.0
1.88
2.12
1.0
0.87
1.13
(31) Position controls as listed in (a) through (c) below:
(a) TIME/DIV OR DLY TIME switch to 20 ns.
(b) DLY'D time/division switch to 2 ns PUSH FOR DLY'D.
(c) MAIN TRIGGERING LEVEL control for stable display.
(32) Repeat (25) through (29) above.
(34) Position controls as listed in (a) through (c) below:
(a) TIME/DIV OR DLY TIME switch to 10 ns.
(b) DLY'D time/division switch to 1 ns PUSH FOR DLY'D.
(c) MAIN TRIGGERING LEVEL control for stable display.
(35) Repeat (25) through (29) above, except omit DELAY TIME MULT dial setting
check at 1.0.
b. Adjustments
(1) Deenergize oscilloscope and connect TI to oscilloscope horizontal compartment
using extender. Energize oscilloscope.
(2) Position controls as listed in (a) through (c) below:
(a) TIME/DIV OR DLY TIME switch to 1 ms.
(b) DLY'D time/division switch to 1 ms PUSH FOR DLY'D.
(c) DELAY TIME MULT dial to 0.0.
(4) Adjust MAIN TRIGGERING LEVEL control for a stable display and
POSITION control to aline 1st time marker with 1st vertical graticule line (extreme left)
on oscilloscope.
(5) Adjust SWP CAL control (front panel) for 1 time marker per division over center
8 divisions.
PIN: 046203-001