EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287 or AN/GSM-705. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Adapter, SMC jack to BNC plug, Hewlett-Packard Model 1250-0831; Extender
Boards, 10-terminal circuit board extenders (05345-60201 and 05345-60202); and Extender
Cable, Hewlett-Packard Model 05345-60205.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: 105 to 125 V ac
General Radio, Type
Accuracy: 1%
W10MT3AS3 (7910809) or
Ridge, Model 9020A (9020A) or
Ridge, Model 9020F (9020F)
(SG-1288/G)
Range: 10 Hz to 30 MHz
Amplitude: 0 to 40 mV rms
Accuracy: 3% (1.5 dB)
Range: 10 MHz
Tracor, Model 527E (527E)
METER
Resolution: 1 part in 10-10 per day
Range: -2 to +2 Vdc
Hewlett-Packard, Model
Accuracy: 0.000125%
3458A (3458A)
Range: 0 to 15 V dc
Accuracy: 0.03%
(OS-303/G)
Sensitivity: 50 mV/div
Sweep rate: 10 ns/div
Accuracy: 3%
(SG-1207/U)
Range: 30 to 500 MHz
Amplitude: 0 to 40 mV rms
Accuracy: 3% (1.5 dB)
Datum, Model ET6000-75
TIME/FREQUENCY
(13589305)
WORKSTATION
Accuracy: 1.25 parts in 10
-10 per day