TB 9-6625-1997-35
Common name
Description (part number)
CABLE No. 1 1
36-in., RG-223/U, BNC PLUG TERMINATION
CABLE No. 2 1
36-in., RG-62/U, BNC PLUG TERMINATION
BNC plug to jack (50 to 25Ω) TERMINATIONS (CN 900/CU)
ADAPTER No. 1
BNC plug to jack (100 to 50Ω) terminations CN-901/U)
ADAPTER No. 2
BNC plug (100 to 50Ω) to jack terminations (CN-9f02/U
ADAPTER No. 3
1
Furnished with TI.
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TMs 11-6625-586-12 and 11-6625-586-45.
d. When
indications specified in
paragraphs 8 through
15 are not within
tolerance,
perform the power supply check prior to making adjustments. After adjustments are
made, repeat paragraphs 8 through 15. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all control and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance
of this calibration. DEATH ON CONTACT may result if
personnel fail to observe safety precautions. REDUCE
OUPUT(S) to minimum after each step within the
performance check where applicable.