TB 9-6625-1998-35
(4) Reposition frequency dial to indicate 200.
NOTE
Remove frequency dial knob and loosen set screws to reposition
dial to 200. Tighten set screws after repositioning dial.
(5) Set frequency dial in turn to 100, 60, 40, and 20. If audio analyzer does not
indicate within 2 percent of settings, repeat (1) through (4) above for best in-tolerance
condition.
(6) Repeat a(4)above. If audio analyzer does not indicate within limits specified,
perform (7) through (15) below.
(7) Set C14 (fig. 1) to the half-open position.
(8) Set FREQ RANGE MULTIPLIER switch to X1K and frequency dial to 20.
(9) Adjust C5 and C13 (fig. 1) until audio analyzer indicates 20.000 kHz (R).
(10) Turn OUTPUT CONTROL fully cw.
(11) If required, readjust C5 and C13 (fig. 1) until OUTPUT RMS meter indicates at
least 1.15 and audio analyzer indicates 20.000 kHz.
NOTE
(12) Set frequency dial to 200.
(13) Adjust C14 (fig. 1) until audio analyzer indicates 200.00 kHz (R).
(14) Set frequency dial to 20 and repeat (11) through (13) above for best in-tolerance
condition.
(15) Repeat a(1) through (4) above and, if required, (1) through (14) above for best
intolerance condition.
(16) Set FREQ RANGE MULTIPLIER switch to X100 and frequency dial to 20.
(17) Readjust frequency dial until audio analyzer indicates 2000.0 Hz.
(18) Reposition frequency dial to indicate 20.
9