TB 9-6625-2013-50
9. Resistance Decade Dials
X.0001Ω
, X.01Ω
, X0.1Ω
, and X1Ω
Accuracy
a. Performance Check
(1) Set TI decade dials to settings listed in table 3. Multimeter indications will be
Test instrument
decade dial settings1
(Ω)
X1
X0.1
X.01
X.0001
Min
Max
0
1
1
00
0.1090
0.1110
0
2
2
00
0.2190
0.2210
0
3
3
00
0.3290
0.3310
0
4
4
00
0.4390
0.4410
0
5
5
00
0.5490
0.5510
0
6
6
00
0.6590
0.6610
0
7
7
00
0.7690
0.7710
0
8
8
00
0.8790
0.8810
0
9
9
00
0.9890
0.9910
0
X
X
100
1.1090
1.1110
0
X
11
100
1.1190
1.1210
1
9
X
00
1.9989
2.0011
2
9
X
00
2.9988
3.0012
3
9
X
00
3.9987
4.0013
4
9
X
00
4.9986
5.0014
5
9
X
00
5.9985
6.0015
6
9
X
00
6.9984
7.0016
7
9
X
00
7.9983
8.0017
8
9
X
00
8.9982
9.0018
9
9
X
00
9.9981
10.0019
X
9
X
00
10.9980
11.0020
1TI
decade dials not listed are set to zero.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Set TI decade dials to 000010.0100. If multimeter indication is not between
10.0087Ω and 10.0113Ω, perform b below.
(2) Repeat technique of (1) above for TI decade dial settings and multimeter
indications listed in table 4. If multimeter indications are not within limits specified at
each setting, perform corresponding adjustment listed in table 4.