TB 9-6625-2013-50
b. Adjustments.
Adjust X1K
R1 (fig. 1)
for a
1.000010 kΩ
indication on
multimeter (R).
a. Performance Check
(1) Set TI decade dials to 010000.0100. If multimeter indication is not between
9.999759 and 10.000261 kΩ, perform b below.
(2) Repeat technique of (1) above for TI decade dial settings and multimeter
indications listed in table 7. If multimeter indications are not within limits specified at
each setting, perform corresponding adjustment listed in table 7.
Test instrument
Multimeter indications
Adjustments for X10K
decade dial
(kΩ)
settings
Min
Max
(R)
020000.0100
19.999509
20.000511
R2
030000.0100
29.999259
30.000761
R3
040000.0100
39.999009
40.001011
R4
050000.0100
49.998759
50.001261
R5
060000.0100
59.998509
60.001511
R6
070000.0100
69.998259
70.001761
R7
080000.0100
79.998009
80.002011
R8
090000.0100
89.997759
90.002261
R9
0X0000.0100
99.997509
100.002511
R10
Adjust resistor X10K R1 (fig. 1) for a 10.000010 kΩ multimeter
b. Adjustments.
indication (R).
14. X100 kΩ
Decade Dial Accuracy
a. Performance Check
(1) Set TI decade dials to 100000.0100. If multimeter indication is not between
99.99751 and 100.00251 kΩ, perform b below.
(2) Repeat technique of (1) above for TI decade dial settings and multimeter
indications listed in table 8. If multimeter indications are not within limits specified at
each setting, perform corresponding adjustment listed in table 8.