TB 9-6625-2020-24
(10) Set calibrator for a 3.0984 V, 100 Hz output. Adjust R1024 (fig. 2) until TI
indication just begins to flash (starts to blink).
(11) Set calibrator for a 23.238 mV, 100 Hz output. Adjust R1012 (fig. 2) until TI
indication just begins to flash (starts to blink).
a. Performance Check
NOTE
Do not perform power supply check if all other parameters are
within tolerance.
and LOW.
(1) Short TI VOLTS/
(2) Connect multimeter INPUT HI to TP1542 (fig. 2) and LO to TI chassis ground.
If multimeter does not indicate between +15.74 and +15.76 V dc, perform b below.
b. Adjustments. Adjust R1771 (fig. 2) until multimeter indicates +15.75 V dc. (R)
a. Deenergize and disconnect all equipment
b. Annotate and affix DA label/form in accordance with TB 750-25.