TB 9-6625-2024-35
(5) Adjust C594 (fig. 4) to align 1 marker on each vertical graticule line over
center 8 divisions (R).
(6) Adjust C681 (C691 for Tektronix 7B53A, 7B53AN, and TD-1159/U) (fig. 4) to
align 1 marker on each vertical graticule line over center 8 divisions (R).
22. Delay Time Multiplier
Start/Stop, Linearity, and Time Accuracy
a. Performance Check
(1) Position TI controls as listed in (a) through (c) below:
(a) TIME/DIV switch to 1 mS.
(b) DLY TIME switch to 10 S and PUSH FOR DLY'D.
(c) DELAY TIME MULT dial to 1.0.
(3) Adjust horizontal POSITION control to align second marker with center
vertical graticule line.