TB 9-6625-2057-35
Paragraph
Page
Frequency accuracy and stability .......................
Frequency response..............................................
Distortion ..............................................................
Final procedure ....................................................
CALIBRATION PROCESS FOR SIGNAL
HEWLETT-PACKARD, MODEL 204B
Preliminary instructions......................................
Equipment setup ..................................................
Frequency accuracy and stability .......................
Output voltage......................................................
Frequency response..............................................
Distortion ..............................................................
Power supply ........................................................
Final procedure ....................................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
Packard, Models 204B, 204C, and H20-204C. The manufacturer's manuals, TM 11-6625-
602-40-1, TM 11-6625-602-45, TM 11-6625-1589-15 and TM 11-6625-1589-15-1 were used
as the prime data sources in compiling these instructions. The equipment being calibrated
will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in the text.
b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.