TB 9-6625-2059-24
TI multimeter
RANGE switch positions
OHMS scale indications
Min
Max
30
28.5
31.5
X1 Ω
X10 Ω
30
285
315
X100 Ω
30
2850
3150
X1K Ω1
30
28.5 k
31.5 k
X10K Ω1
30
285
k
315 k
1 If
required, readjust A2A1A1R123 (fig. 8) for best in-tolerance indication.
(13) Set TI multimeter FUNCTION switch to HI-Z and RANGE switch to 1V.
(14) Connect calibrator to TI multimeter INPUT.
(15) Adjust calibrator frequency for 1 kHz frequency and output for a 1 V indication
on TI multimeter. If calibrator does not indicate within limits specified in first row of
(16) Repeat technique of (15) above for remaining settings and indications listed in
25