TB 9-6625-2059-35
Test instrument parameters
Performance specifications
Frequency
Range: 100 Hz to 999.9999 MHz1
Accuracy: 0.00005% (10 to 999 MHz)
5 Hz (1 to 10 MHz)
Residual FM
<100 Hz peak
Output level
V x 100
Range: -35 to -90 dBm
Accuracy: 2.5 dB to 199.9999 MHz
4.0 dB from 200 to 399.9999 MHz
6.0 dB from 400 MHz and above
HI LVL
Range: 0 to -35 dBm relative to 0 dBm indication
Accuracy: 2.5 dB, 20 kHz to 199.9999 MHz
4.0 dB from 200 to 399.9999 MHz
6.0 dB from 400 MHz and above
Norm2
Range: -75 to -120 dBm
Accuracy: 2.5 dB to 199.9999 MHz
4.0 dB from 200 to 399.9999 MHz
6.0 dB from 400 MHz and above
Power monitor:
Frequency
Range: 1 to 1000 MHz
Power
0 to 4, 0 to 40, and 0 to 400 W
Accuracy: 7% of reading. 3% FS from 1 to 600 MHz
20% of reading 3% FS from 600 to 1000 MHz
External vertical input
10 and 100 mV/div; 1 and 10 V/div
Horizontal sweep
Accuracy: 10%
10 and 1 ms/div; 100 and 10 s/div
Frequency:
Variable tone
Range: 10 to 20 kHz
Accuracy: 0.01%
Fixed tone
Range: 1 kHz
Accuracy: 20 Hz
0 to 2.5 V rms minimum either tone into 150 Ω
Output level
Distortion
<1.5% 10 to 100 Hz
<0.7%, 100 Hz to 20 kHz
<2%, 1 kHz fixed tone
Spectrum analyzer:
Dynamic range
Range: 70 dB from -30 to -100 dBm
Accuracy: 4 dB relative to -50 dBm
Continuous from 0.5 to 5 MHz from center frequency
(1 to 10 MHz span)
2.0 V above 1 MHz
Sensitivity
Range: 1.5 kHz, 5 kHz, 15 kHz FS
Accuracy: 0.5 kHz on 15 kHz range
0.2 kHz on 5 kHz range
0.1 kHz on 1.5 kHz range
Deviation kHz meter
Range: 2, 6, and 20 kHz ranges
Accuracy: 4% reading 3% FS
See footnotes at end of table.
3