TB 9-6625-2059-35
TI multimeter
Calibrator indications (Vdc)
RANGE
Meter indications
switch settings
0 to 3 scale
0 to 10 scale
Min
Max
.1
---
10
0.097
0.103
.3
3
---
0.291
0.309
3
3
---
2.91
3.09
10
---
10
9.7
10.3
10
---
8
7.7
8.3
10
---
6
5.7
6.3
10
---
4
3.7
4.3
10
---
2
1.7
2.3
101
---
10
97
103
30
3
---
29.1
30.9
100
---
10
97
103
300
3
---
291
309
102 , 3
---
10
9.7
10.3
1Set
probe X1-X10 switch to X10 for this check only.
2Set FUNCTION switch to DC - (see CAUTION below) and reverse input at calibrator.
CAUTION
Before performing negative dc measurement, ensure that dc output of calibrator is
floating from chassis ground. Failure to do so may result in serious damage to
3If out-of-tolerance, adjust calibrator for 10.0000 and adjust A2A1A1R82 (fig. 8) for 10 V indication on multimeter. (R)
(9) Set TI multimeter FUNCTION switch to OHMS and RANGE switch to X1 Ω.
If meter does not indicate (infinity), perform b (4) below.
(10) Connect resistance standard to TI multimeter INPUT, using probe in table 3
(set to X1).
29