TB 9-6625-2074-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AC DIVIDER
Range input: 1 to 10 V ac at 10 kHz
Fluke, Model 7405A-4207 (7405A-
4207)
Output voltage: 1 PV to 10 mV ac
Combined accuracy with calibrator: 0.5%
CALIBRATOR
Frequency range: 10 kHz
Fluke, Model 5720A (5700A/EP)
Voltage range: 97 to 3090 mV rms
(p/o MIS-35947); w amplifier,
Fluke 5725A/AR (5725A/AR))
Stand alone accuracy: r0.5%
Accuracy with ac divider: r0.5%
MEASURING RECEIVER
Measuring Receiver system
mV range: 100 mV rms
N5530S consisting of: Spectrum
Accuracy: 0.5%
Analyzer Agilent, Model E4440A
(E4440A), Power Meter Agilent,
Model E4419B (E4419B), and
Sensor Module Agilent, Model 504
(504)
Range: -15.1 to +15.1 V dc
Hewlett-Packard, Model 3458A
Accuracy: 0.17%
(3458A)
POWER SPLITTER
Weinschel, Model 1870A
Port-to-port tracking accuracy: 0.15 dB
(7916839)
Range: 100 kHz to 1.2 GHz
Aeroflex, Model 2023B (2023B) or
Output amplitude: 100 mV
(SG-1207/U)
Common name
Description (part number)
Adapter
50: BNC adapter Boonton, Model 91-8B (p/o TI)
Adapter
Boonton, Model 91-14A (p/o TI)
Probe tip
Boonton, Model 91-13B (p/o TI)
Termination
Boonton, Model 91-15A (p/o TI)