TB 9-6625-2074-24
(11) Repeat technique of (10) above for remaining TI pushbutton settings in table 13.
Calibrator output
TI pushbutton
TI indication
(V ac)
TI adjustment
setting
(mV)
Min
Nominal
Max
b(3)
30
mV
0.0288
0.030
0.0312
30
b(4)
100
mV
0.096
0.100
0.104
100
b(5)
300
mV
0.291
0.300
0.309
300
b(6)
1000
mV
0.970
1.0
1.030
1000
b(7)
3000
mV
2.910
3.0
3.090
3000
(12) Set calibrator to standby.
(13) Set TI pushbuttons as listed in table 14.
(14) Adjust calibrator controls for 1MHz and a full-scale indication at each TI
pushbutton setting. Calibrator will indicate within limits specified.
Calibrator output
TI pushbutton
TI indication
(V ac)
setting
(mV)
Min
Nominal
Max
1000mV
0.97
1.0
1.03
1000
3000mV
2.91
3.0
3.09
3000
(15) Disconnect TI probe from calibrator.
(17) Zero and calibrate the measuring receiver power sensor module.
(18) Set measuring receiver to measure RF power in mV.
(19) Adjust signal generator frequency controls to frequency setting listed in table 15.
20