TB 9-6625-2083-24
NOTE
If standardizer does not have sufficient range to obtain
optimum square wave, adjust C9 (fig. 6) to midrange and
readjust standardizer (R).
Figure 7. Bottom and left side view - adjustment locations.
(4) Repeat technique of (3) above for VOLTS/DIV switch settings and adjustments
listed in table 9, but do not readjust standardizer.
(5) Set oscilloscope calibrator for a CHAN 2, EDGE mode output of 100 mV at 1 kHz.
(6) Set TI MODE switch to CH 2 and repeat technique of (3) and (4) above, using
CH 2 controls and adjustments listed in table 9.
Test instrument
Adjustments (R)
VOLTS/DIV
Square corner
Flat top
switch
CH 1
CH 2
CH 1
CH 2
settings
0.1
---
---
C9
C109
0.2
C5D
C105D
C5A
C105A
21
C6E
C106E
C6A
C106A
1Remove
termination.
25