• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
B Sweep Calibration
Magnified Sweep

TB-9-6625-2083-35 Oscilloscope AN/USM-273 (OS-188U and Tektronix Type 453) and Tektronix Types 453A 453A3 453MOD703H 454 454MOD163D AND 468 Manual
Page Navigation
  23    24    25    26    27  28  29    30    31    32    33  
TB 9-6625-2083-35
Table 11. Sweep Timing Accuracy
Test instrument
Oscilloscope calibrator
A AND B TIME/DIV AND
MARKER mode
Err display
DELAY TIME switch
output
indication limits (%)
settings
.1 s
.1 s
3
.2 s
.2 s
3
.5 s
.5 s
3
1
1
1  s
1  s
3
2
2
2  s
2  s
3
5  s
5  s
3
10  s
10  s
3
20  s
20  s
3
50  s
50  s
3
.1 ms
.1 ms
3
.2 ms
.2 ms
3
.5 ms
.5 ms
3
1  ms
1  ms
3
2  ms
2  ms
3
5  ms
5  ms
3
10  ms
10  ms
3
3
3
20  ms
20  ms
3
50  ms
50  ms
3
.1 s
.1 s
3
.2 s
.2 s
3
.5 s
.5 s
3
1  s4
1  s4
3
2  s
2  s
3
4
4
5  s
5  s
3
4
4
1Adjust
C740A (fig. 8) for A TIME/DIV and C940A (fig. 8) for B TIME/DIV AND DELAY TIME (R).
2Adjust
C740C (fig. 8) for A TIME/DIV and C940 C (fig. 8) for B TIME/DIV AND DELAY TIME (R).
3Set A SWEEP MODE switch to NORM TRIG.
4A SWEEP only.
b. Adjustments. Perform adjustments listed in table 11 footnotes.
28. High Speed Linearity
a. Performance Check
(1) Position switches as listed in (a) through (c) below:
(a) CH 1 VOLTS/DIV to 50 mV.
(b) A AND B TIME/DIV AND DELAY TIME to .05 s.
(c) HORIZ DISPLAY to A.
(2) Set oscilloscope calibrator for a CHAN 1, MARKER mode output of 5 ns.
(3) Position MAG switch to X10.
(4) Position second marker at second graticule line so that sweep starts at first
graticule.
(5) If display does not have optimum linearity (even spacing) over center 8 divisions
of crt, perform b (1) below.
30






Western Governors University

Privacy Statement
Press Release