Figure 12. Interface board.
Storage Timing

TB-9-6625-2083-35 Oscilloscope AN/USM-273 (OS-188U and Tektronix Type 453) and Tektronix Types 453A 453A3 453MOD703H 454 454MOD163D AND 468 Manual
Page Navigation
  39    40    41    42    43  44  45    46    47    48  
TB 9-6625-2083-35
Figure 13. Timing board.
(5) Adjust R578 MAG REGISTRATION (fig. 12) to position time marker to center
vertical graticule line (R).
(6) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for
an Err display of 0.0%. Adjust R169 B SWP CAL (fig. 13) for 1 marker per division (R).
(7) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for
an Err display of 0.0%. Adjust C120 10 S TIMING (fig. 13) for 1 time marker per division (R).
(8) Adjust C120 10 S TIMING (fig. 13) to align displayed time marker with 1st
vertical graticule line. Repeat a (10) through (12) above until no further improvement is
noted (R).
(9) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for
an Err display of 0.0%. Adjust C126 A SWP HI SPEED TIMING (fig. 13) for 1 marker per
division (R).
(10) Adjust C126 A SWP HI SPEED TIMING (fig. 13) to align displayed time marker
with center vertical graticule line.  Repeat performance check and adjustment a (16)
through (19) above until no further improvement is noted (R).
(11) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for
an Err display of 0.0%. Adjust C159 B SWP HI SPEED TIMING (fig. 13) for 1 time marker
per division (R).