TB 9-6625-2084-35
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of
Equipment Required
Common name
Minimum use
Manufacturer and model
(official nomenclature)
specifications
(part number)
CALIBRATION FIXTURE
Range: 1 kHz to 1 MHz
Tektronix, Type 067-0587-02
(STANDARDIZER)
Risetime:
0.875 ns
(067-0587-02)
Voltage output:
John Fluke, Model 5820A-5C-
CALIBRATOR
Range:
50 mV to 5 V at 1 GHZ (5820A-5C-GHZ)
kHz
Accuracy:
0.25%
Time marker:
Range:
5 ns to 1 ms
3%
Accuracy:
Voltage input:
Range:
1520 to +135 V dc
0.25%
Accuracy:
Must be compatible w/test
Tektronix, Type 7B Series
1
TIME BASE
instrument and vertical
amplifier
Must be compatible w/test
Tektronix, Type 7A Series
1
VERTICAL AMPLIFIER
instrument and time base
1
Provided with TI.
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire procedure before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test, and whenever the test
requirement is not met, take corrective action before continuing with this calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. When indications specified in
paragraphs 8 through 13 are not within
tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 13. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.