TB 9-6625-2097-35
NOTE
Unless otherwise specified, verify the result of each test and,
whenever the test requirement is not met, take corrective
action before continuing with the calibration. Adjustments
required to calibrate the TI are included in this procedure.
Additional maintenance information is contained in the
manufacturer's manual for this TI.
NOTE
Unless otherwise specified, all controls and control settings
refer to the TI.
13. Equipment Setup
a. Remove protective cover from TI.
b. Slide all manual ink valves with tubing attached to ON position on recorder (A6).
c. Connect TI to recorder, using extender (B3).
d. Set front panel controls as listed in (1) through (7) below:
(1) FULL SCALE switch to 1.
(2) ZERO SUPPRESSION switch OFF.
(3) ZERO SUPPRESSION control ccw to 0.
(4) FILTER switch OFF.
(5) MODE switch to X1 VOLTS.
(6) SENSITIVITY control fully cw.
(7) RANGE switch to X1.
e. Set DC INPUT
switch S-201 to left
(fig. 3).
f. Energize equipment and allow 15 minutes for warmup and stabilization.
9