TB 9-6625-2114-24
Table 2. Minimum Specifications of Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
CALIBRATOR
Range: 0 to 315 V ac at 400 Hz
Fluke, Model 5720A (5720A) (p/o
891 to 909 V ac at 10 kHz
MIS-35947); w/ amplifier, Fluke,
Accuracy: 0.75%
Model 5725A/R (5725A/R)
Range: 0.3 V ac; 20 Hz to 1 MHz
Accuracy: 0.5%
Range: 0 to 1030 V dc
Accuracy: 0.5%
Range: 0 to 150 MA dc
Accuracy: 0.75%
MEASURING RECEIVER
Measuring receiver system N5530S
Amplitude: 0.3 V
consisting of: Spectrum Analyzer,
Agilent Model E4440A (E4440A),
Accuracy:
1.0% 60 to 90 MHz
Power meter, Agilent Model
4.5% 150 to 700 MHz
E4419B (E4419B), and Sensor
module, Agilent Model N5532A opt.
504 (504), 518 (518), 526 (526)
Range: 4.9 to 6.4 V ac
Hewlett Packard, Model 3458A
0 to 51.5 mV dc
(3458A)
Accuracy: 0.75%
(SG-1207/U)
Amplitude: 0.3 V
A
Biddle-Grey, Model 71-631
Range: 0 to 1.1 M
(7910328)
ccuracy: 1.25%
A
Beckman, Model CR10M (8598965)
Range: 0 to 10 M
ccuracy: 1.25%
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the applicable sections before beginning
the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals and TM 11-6625-200-15
for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.