(18) Adjust signal generator amplitude for a .3 V ac indication on TI. Measuring
receiver will indicate within 1 dB.
(19) Repeat (16) through (18) above at 90, 150 and 250 MHz.
(20) Repeat technique of (16) through (18) above at 350, 550 and 700 MHz.
Measuring receiver will indicate within 3 dB.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Set SELECTOR switch to OHMS.
(2) Adjust OHMS ADJ for ∞ meter indication.
(3) Short OHMS and COMMON leads together and adjust ZERO ADJ for 0
meter indication.
(4) Separate OHMS and COMMON leads.
(5) Repeat (2) through (4) above until no further adjustment is necessary.
(6) Connect OHMS and COMMON leads to resistance standard No. 1.
(7) Set resistance standard for a 10Ω indication on TI. If resistance standard
setting is not between 9 and 11Ω, perform b below.
(8) Disconnect TI from resistance standard.
(9) Repeat technique of (2) through (8) above using settings and indications in
table 5. Resistance standard settings will be within limits
specified in
table 5.
Resistance standard
Test instrument
indications
RANGE
Indications
switch settings
(OHMS scale)
Min
Max
Ω
Ω
RX10 3V
10
95
105
Ω
Ω
RX100 10V
10
950
1050
kΩ
10.5 kΩ
RX1K 30V
10
9.5
kΩ
kΩ
RX10K 100V
10
95
105
MΩ
1.05 MΩ
RX100K 300V
10
.95
MΩ
10.5 MΩ
RX1M 1000V
dc only1
10
9.5
1Connect
resistance standard No. 2 in series with resistance standard No. 1.
b. Adjustments
(1) Remove OHMS lead from resistance standard and set RANGE switch to
RX10 3V. Adjust OHM ADJ for ∞ meter indication.
Adjust R49 RX1 (fig. 1) for ∞ meter
(2) Set RANGE switch to RX1 1V.
indication (R).
(3) Repeat a (2) through (7) above.