TB 9-6625-2114-35
(10) Connect measuring receiver sensor module input to calibrator OUTPUT HI
and LO.
(11) Set calibrator to OPERATE.
(12) Set measuring receiver to measure RF power in dBm at 1 MHz. Record
measuring receiver indication.
(13) Set calibrator to STANDBY and disconnect from measuring receiver.
measuring receiver indication as close as possible to indication recorded in (12) above.
Press measuring receiver MEASUREMENT DISPLAY RATIO key.
(16) Set signal generator frequency to 60 MHz.
(18) Adjust signal generator amplitude for a .3 V ac indication on TI. Measuring
receiver will indicate within 0.35 dB (4%).
(19) Repeat (16) through (18) above at 90 MHz.
(20) Repeat technique of (16) through (18) above at 250, 500 and 700 MHz.
Measuring receiver will indicate within 1.5 dB.
b. Adjustments. No adjustments can be made.
15